Running LVShim
The objective of the LVShim tool is to fine-tune magnet homogeneity.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 30 minutes | 10 minutes |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| LVShim Phantom Inner Sphere | 1 | 2274476 | - |
| Universal Phantom Pad | 1 | 46-328383P1 | - |
| Required conditions |
|---|
| Confirm that passive shimming is complete and is within specification. |
| Remove all coils from the cradle and phantoms from the bore. |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
Note: For systems running PX28.0 or earlier, use the LV phantom side lobes. For systems running PX29.0 or later, use the LVShim Phantom Inner Sphere with the Universal Phantom Pad. If LV phantom side lobes are used on a system running PX29.0 or later, an error message displays.