- Discovery MR750w and SIGNA™ Architect T 3.0T System Service Methods
- 5690002-2EN Revision 4
- 00000018WIA3058E030GYZ
- id_123737381.5
- Feb 1, 2022 9:27:08 PM
Hydraulic Fluid Draining and Rotary Pump Replacement Videos
Visit the MR Service Documentation Website to view video clips showing the process to replace the rotary pump and drain the table hydraulic fluid.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | 5 minutes | 120 mins. for Pump replacement; 90 mins. for Hydraulic fluid replacement; 15 mins. for Air bubble extraction | 5 minutes |
| Tools and test equipment | ||||
|---|---|---|---|---|
| Item | Quantity | Effectivity | Part number | Manufacturer |
| Hydraulic Fluid Container. Maximum height 3.5 inches (90 mm), must hold 64 oz. (2 liters). Video example is 2 in. high x 6 in. wide x 12 in. long. | 1 | - | - | - |
| Consumables | ||||
|---|---|---|---|---|
| Item | Quantity | Effectivity | Part number | Manufacturer |
| Hydraulic Fluid | 2 qts | - |
See FRU manual | - |
| Absorbent Towels | As Needed | - |
N/A | - |
| Wire Ties | As Needed | - |
N/A | - |
| Replacement parts | ||||
|---|---|---|---|---|
| Item | Quantity | Effectivity | Part number | Manufacturer |
| Rotary Pump | 1 | - |
See FRU manual | - |
| ||||||||
About this task
Overview
Visit the MR Service Documentation Website to view three different video clips showing the process to:
-
Replace the rotary pump
-
Drain the table hydraulic fluid
-
Remove air bubbles
Note:
These videos are for reference only and are not intended to replace the existing service manual procedures. Refer to Patient Table Hydraulic Rotary Pump Replacement and Hydraulic Fluid Level Check and Fill for detailed instructions.
Procedure
Finalization
No finalization steps.