Body Hybrid Replacement

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
2 for MR750; 1 for MR450, MR450w, MR450w GEMNot Applicable30 minutes30 minutes
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Non-Magnetic Service Tool Kit1-

5112581

-
Table 4. Safety
Topic ID: id_SL1901440-1057647

Removing Body Hybrid Splitter

Procedure

  1. Perform LOTO on the PDM-RF power for the system. See the MR Service Safety Manual, PN 5452735.
  2. Make sure the scanner is idle. Select End Exam at any point in the scan process.
  3. Remove the module from the magnet room, and keep it as far away from the magnet as possible.
Topic ID: id_SL1901450-1057647

Installing Body Hybrid Module

Procedure

  1. Move the module into the magnet room, and keep it as far away from the magnet as possible.
  2. Replace all covers in reverse order.
  3. Remove LOTO and reapply power to the system. See the latest revision of the MR Service Safety Manual, PN 5452735.