Installing software options with eLicense generated keys
This procedure describes how to use eLicenses to generate and install software options on MR scanners.
About this task
The MR system must be tied to the Global Order Number (GON) through the Host ID and System ID of the scanner. After the site information is entered, the keys can be generated and distributed by e-mail, downloading, saving to CD/DVD, or printing. This file can then be installed on the system through CD/DVD, InSite auto download, SFTP, or manual entry.
The following tasks are included in this procedure:
For additional information on how to generate and obtain these keys from the eLicense server, see the eLicense User Guide under the HELP tab of the eLicense website.