Leak Sensor Location and Function

There are two Leak sensor routes.

Note:

Leak Sensor 1 is not used for Optima MR360/Brivo MR355 system

  • Leak Sensor 2

  • Leak Sensor 3

When leak sensor detects the leakage, cabinet monitor will react as follows.

  • If the Leak Sensor2 detects the leakage, LK2 LED is lit and error message is displayed on Host and Emergency Stop will work.

  • If the Leak Sensor3 detects the leakage, LK3 LED is lit and error message is displayed.

Figure 1. leak sensor routing