- id_12373940
- Version: 1.12
- Date: Dec 16, 2019 2:22:43 PM
DTRSW and Preamp Combiner Replacement
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | Not Applicable | 30 minutes | 30 minutes |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Non-magnetic Phillips Screwdriver | 1 | - | - | - |
| Basic Hand Tool Set | 1 | - | - | - |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Dual Drive Transmit Receive Switch | 1 | - |
See FRU Manual |
- |
| Preamp Combiner | 1 | - |
See FRU Manual |
- |
|
Overview
This procedure includes the following FRU replacement procedures:
-
Dual drive transmit receive switch (DTRSW). Note that the DTRSW for MR750 is different from other products like MR750w and Signa Architect. Refer to the FRU Manual for the correct part number.
-
Preamp combiner
Removing DTRSW or Preamp Combiner
Procedure
- Perform LOTO on the PDM-RF power for the system. See the MR Service Safety Manual, PN 5452735.
- Disconnect the cables connected to the DTRSW, as follows:
-
If you replace the DTRSW, remove all cables.
-
If you replace the preamp combiner, remove cables from J9, J10, J11, and the Hart ID connector on preamp combiner.
Figure 1. DTRSW Cable Connections (RRx)
-
- To replace a defective preamp combiner, remove the 4 screws
attaching it to the DTRSW.
- If you are replacing the DTRSW, remove the preamp combiner first
as shown in Step 3. Remove the 4 screws holding the DTRSW to the outside of the magnet.
- If you are replacing the DTRSW, remove the preamp combiner and reinstall it on the new DTRSW.
caution- While carrying any module out of the magnet room, stay as far away from the magnet as possible.
Installing DTRSW or Preamp Combiner
Procedure
- While carrying any module into the magnet room, stay as far away from the magnet as possible.
caution- Secure the module, as follows:
-
For the DTRSW, secure the module to the magnet by tightening the four screws.
-
For the preamp combiner, secure the module to the DTRSW by tightening the four nuts.
-
- Connect all cables that were previously removed. See Figure 1.
Finalization
Procedure
- Replace the side covers.
- Remove LOTO from the PDM-RF power for the system. See MR Service Safety Manual, PN 5452735.
- Perform TR Dynamic Disable Calibration.
- If hardware was replaced, perform System gain calibration for body.
- Perform a body SNR test.
- Perform SaveInfo to save new calibration values.
- Complete a body and head test scan to make sure the system is functional.