Checking the GOC battery replacement
Check the battery replacement sticker on the GOC.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 5 minutes | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
Procedure
Check the battery replacement sticker on the GOC.
The sticker should be attached to lower left or right side of the GOC. Batteries should be replaced every three years.
- If the date on the battery replacement sticker is within 12 months, order battery and replace during the next PM per the Lower Level FRU Battery Replacement procedure for that computer model.
- If the GOC does not have a sticker, order a battery and replace during the next PM per the Lower Level FRU Battery Replacement procedure for that computer model.