Replacing the PS filter box

Replace the PS filter box in the Integrated System Cabinet (ISC) Penetration Wall (PW).

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-30 minutes15 minutes
Tools and test equipment
ItemQuantityPart numberManufacturer
Digital Voltmeter (DVM)146-194427P284-
Red Warning LOTO Tag246-194427P322-
Multi-Locking Device (if multiple service personnel are involved)146-194427P313-
Red LOTO Personal Lock Wrap22393068-
Brass Master Padlock with Brass Shackle246-194427P320-
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Replacement parts
ItemQuantityPart numberManufacturer
ISC Penetration Wall Filter Box16859100-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.