Important: The safety analyzer must be located as far away from the magnet as possible to successfully execute this test. Getting closer to the magnet affects the safety analyzer and alters the tool’s ability to take an accurate measurement.
Procedure
Plug the analyzer into a power source in the magnet room which allows access to the patient monitor accessory hook on the left side of the magnet enclosure, while staying at least 20 inches (51 cm) from the magnet.
The patient monitor accessory hook is on the left side of the magnet, regardless of system version. The example shown below is an MR450 system.
Place the L2 (Neutral) switch in the CLOSED position.
Figure 5. Dale 601/601E Safety Analyzer
Note: Be sure to pause in the OFF (middle) position when switching polarity from normal to reverse.
Place the Outlet (Polarity) switch in the NORMAL position.
On the Dale 601/601E, set the Function switch to PATIENT LEAD LEAKAGE.
Move the lead switch through the lead positions shown in Table 1. Verify that the leakage current displayed on the meter is less than 50 microamps and record the values on the data sheet.
Note: Typically, the Patient Lead Leakage (lead-to-ground) and the Patient Aux Current (lead-to-lead) readings are close to zero and less than 50 microamps. Measurements exceeding 50 microamps indicate an issue in the corresponding lead.
Table 1. Lead Positions
Lead select position
RL
RA
LA
LL
On the Dale 601/601E, set the main selector switch to PATIENT AUX CURRENT.
Move the lead switch through the lead positions shown in Table 1. Verify that the leakage current displayed on the meter is less than 50 microamps and record the values in the data sheet.
On the Dale 601/601E, set the main selector switch to M.A.P. (Mains on Applied Parts sink current).
Make sure that the leads are not touching the floor.
Move the lead switch to measure All Leads, which is the sixth position all the way to right on the switch.
On the LIFT/Ground/MAP switch, press and hold M.A.P. Make sure the leakage current displayed on the meter is less than 50 microamps and record the values on the data sheet.
Note: Be sure to pause in the OFF (middle) position when switching polarity from normal to reverse.
Repeat steps Step 7 - Step 14 with the polarity switch in the REVERSED position.