- Discovery MR750 3.0T System Service Methods
- 5690009-2EN Revision 4
- 00000018WIA309FC410GYZ
- id_20044596.1
- Jul 13, 2021 4:10:15 PM
Cleaning and replacing P-port receptacles (ODU connectors)
The P-Port receptacle must be cleaned and lubricated on each of the system P-ports.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 5 minutes per connector to clean and lubricate; 30 minutes to replace | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Latex-Free Nitrile Gloves (8 mil) | 1 pair | 46-194427P400 | - |
| Spindle Tip Replacement Kit, ODU P-Port Plug Tool | 1 | 5150668-33 | - |
| Safety Shoes | 1 pair | - | - |
| Safety Glasses | 1 | - | - |
| 2 mm Allen Wrench | 1 | - | - |
| Consumables | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| P-Port Spindle Lubrication Kit | 1 | 5150668-32 | - |
| HDv P-Port Replacement Spindle Tip Kit | 1 | 5150668-31 | - |
| Cotton Swabs | As required | - | - |
| Soft Cloths | As required | - | - |
| Required conditions |
|---|
| Order the lubricant kit prior to starting this procedure. |
| Safety | ||||
|---|---|---|---|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. | ||||
| ||||
About this task
The P-Port receptacle must be cleaned and lubricated on each of the system P-ports (P1 and P2 on the LPCA and P3 and P4 on the 32-channel table). Also, the coil connector must be cleaned and tested for proper fit to each P-port receptacle.