Planned Maintenance (PM) schedule overview
Planned Maintenance (PM) tasks are developed from design requirement and operations/use assumptions to provide for the minimum required maintenance that maintains specified equipment design performance, high image quality, and safety. GE Healthcare (GEHC) advanced service methods are developed to maximize efficiency and use system access time as productively as possible, as described in the lifecycle maintenance statement.
- The Warranty IB PM Schedule Concept: The Warranty PM Service requirement is different from the lifecycle Out of Warranty (IBOW) PM Service requirement due to manufacturing tests/checks, and installation PM checks, done during the install process, and the newness of the equipment and parts during the first 12 months after install.
- The Remote (Digital) PM Schedule Option: The use of remote, digital performance monitoring in place of on-site performance monitoring in PM tasks and schedules. If the performance tasks are observed digitally, and passed remotely with an approved digital method, then the on-site performance checks are not required for those tasks. Refer to the Digital Remote Check Procedure in Appendix A
- Equipment Operation Check (EOC): EOCs are not PMs, but rather operational checks required by both customer operators and engineers to make sure of safety or equipment performance. EOCs are typically Customer Operator procedures or finalization procedures that an engineer completes if they are present on-site and have performed maintenance. In the case of MR and these products, the Magnet Rundown Unit (MRU) check that is required approximately weekly when the customer or the engineer enters the magnet room will no longer be a PM, but will become an EOC.
PM overview
GE Healthcare equipment is designed and manufactured to be of the highest quality. Likewise, the maintenance requirements in the Customer Operations Manual/User Guides and the Customer Service Manuals are designed to help maintain safe, high quality performance of your equipment for its expected life.
Maintenance and repair work shall be carried out by qualified personnel, trained in maintenance and repair of this medical device in accordance with the manufacturer’s standards within the manufacturing enterprise. Customers should contact their local GE Healthcare Service Sales if they desire to have GE Healthcare complete the PMs with the Advanced (Proprietary) Service Methods and OEM trained service team.
In addition, GEHC systems have required customer, equipment operational checks that the customer should do on a periodic basis to ensure correct system operation or image quality before scanning. The details can be found in the Customer Operator Manuals.
PM scope
The PM tasks, schedules, methods, and procedures apply to the following products and the associated system options (defined in the PM schedules):
- MR 1.5T Discovery 450, Optima 450W, and Optima 450W plus GEM
- MR 1.5T Signa Artist and Signa Artist-T
- MR 3.0T Discovery 750 and Discovery 750W plus GEM
- MR 3.0T Architect and Signa Architect-T