Emergency Stop (E-Stop) planned maintenance
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | - | - |
About this task
Overview
This document provides the procedure for doing a check of the EMERGENCY STOP function of the Power Distribution Unit (PDU) for the Gradient subsystem, and RF subsystem.
Note: To do a check of the EMERGENCY OFF function of the power for the entire system, see Emergency Off (E-Off) planned maintenance.
The PDU EMERGENCY STOP buttons are located on the Operator Workspace Keyboard and on the Magnet Enclosure front cover (two buttons, left and right). A SYSTEM OFF button is located on the System Main Disconnect Panel.
Note: The other EMERGENCY OFF buttons are located by the customer.
Emergency stop
Procedure
Finalization
Finalization
Run the Doing a check scan procedure to make sure the system is functioning normally.