Installs the ICE DCPS into the Power Gradient RF (PGR) cabinet.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 5 minutes | - |
| Tools and test equipment |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
Procedure
- Open the PGR door.
- Disconnect the ICE DC cable from the DCPS.
- Disconnect the AC power cable from the DCPS.
- Loosen the captive fastener securing the DCPS to the tray.
Figure 1. 48V ICE DCPS
| 1 | 48V ICE DCPS tray |
| 2 | 48V ICE DCPS |
| 3 | Captive fastener |
- Remove the ICE DCPS from the ICE DCPS tray.