- Discovery MR750 3.0T System Service Methods
- 5690009-2EN Revision 4
- 00000018WIA30954510GYZ
- id_20045735.3
- Dec 14, 2021 2:33:18 PM
Checking the GOC battery replacement
Check the battery replacement sticker on the GOC.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 5 minutes | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
Procedure
Check the battery replacement sticker on the GOC.
The sticker should be attached to lower left or right side of the GOC. Batteries should be replaced every three years.
- If the date on the battery replacement sticker is within 12 months, order the battery and replace it during the next PM. See Dell T5810 Lower Level FRU Replacement, HP 8400 Lower Level FRU Replacement, HP Z400 Lower Level FRU Replacement, or HP Z420 Lower Level FRU Replacement for your computer model.
- If the GOC does not have a sticker, order a battery and replace it during the next PM. See Dell T5810 Lower Level FRU Replacement, HP 8400 Lower Level FRU Replacement, HP Z400 Lower Level FRU Replacement, or HP Z420 Lower Level FRU Replacement for your computer model.