Front Cover Removal and Installation

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-15 minutes for front cover; 30 minutes for GEM light strips15 minutes
Tools and test equipment
ItemQuantityPart numberManufacturer
Pair: Cut-Resistant Gloves1--
Nonmagnetic Titanium Service Tool Kit, Large Set15112581-
Replacement parts
ItemQuantityPart numberManufacturer
Front Cover2See FRU manual-
Table 1. Safety

About this task

Overview

This document contains the procedures for front cover removal and installation. The front covers consist of two halves and contain the emergency stop and control panels for the system.

Topic ID: id_SL3847945-1095035

Preliminary

Procedure

  1. Turn off and perform LOTO on the RF and PEN cabinets. See the MR Service Safety Manual, PN 5452735.
  2. Undock the patient table and position the table away from the magnet.
Topic ID: id_15367902

Finalization

Procedure