• Object ID: 00000018WIA30B7E030GYZ
  • Topic ID: id_12373721 Version: 2.0
  • Date: Jan 20, 2020 12:54:53 AM

Replacing the hydraulic filter

Removal and installation instructions for replacing the hydraulic filter in the dockable patient table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-40 minutes -
Tools and test equipment
ItemQuantityPart numberManufacturer
3/4 inch Open-End Wrench2--
Ear Syringe1--
Standard Tool Kit with Phillips Screwdrivers1--
Consumables
ItemQuantityPart numberManufacturer
Hydraulic Oil1See FRU Manual-
TowelsAs needed--
Loctite® 5691 bottle (50 ml)46-230614P1-
Replacement parts
ItemQuantityPart numberManufacturer
Hydraulic Oil Filter And Gasket1See FRU Manual-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

About this task

Overview

Do the following procedure as directed to maintain smooth, responsive operation of the patient table hydraulics.