Install the DVD drive in the host computer.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 15 minutes | - |
| Tools and test equipment |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Screwdriver | 1 | - | - |
| ESD Mat | 1 | - | - |
| ESD Strap | 1 | - | - |
| 4 mm Hex Key Wrench | 1 | - | - |
| Replacement parts |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Dell Slimline DVD +/- RW SATA Drive for T5820XL Workstation | 1 | 8780011-101 | - |
| Required conditions |
|---|
| To prevent damage, do not remove the new DVD drive from its antistatic packaging until you are ready to install it. |
|
Use an antistatic wrist strap and mat while performing this procedure to prevent electrostatic discharge (ESD).
|
| Access to both the front and rear of the operator console is required. Move the console away from walls and remove obstacles in the room. |
Procedure
- Slide the DVD drive into the slot on the chassis.
- Tighten the screw to secure the DVD drive to the chassis.
- Install the side cover.
- Hold and align the bottom of the side cover to the chassis.
- Make sure the hook on the bottom of the side cover snaps into the notch on the system.
- Press the system cover until it clicks into place.
Note: The system will not power on without the side cover. Also, the system will shut down if the side cover is removed while the system is on.
- Connect the short ground lead that connects the left side panel to the GOC main chassis at the center of the lead.
- Secure the left side panel to the GOC.
- Get a GE equipment maintenance sticker, and write today's date on the sticker. Put the sticker on the GOC where it is easily visible.
Note: When replacing the hard drive or DVD drive, a maintenance sticker is not required.