- Discovery MR750 3.0T System Service Methods
- 5690009-2EN Revision 4
- 00000018WIA30D79450GYZ
- id_20227211.1
- Jan 21, 2020 3:28:13 PM
Checking the table undock mechanism components
Checks the operation of the table undock mechanism components and primary latch adjustment.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 10 minutes | - |
Procedure
- Move the table into the scan room and dock it.
- Advance the cradle into the bore until the primary latch is exposed.
- Move the cradle back onto the table.
- Undock the table.
- If the cradle fails to move fully off of or onto the table, adjust the cradle home table down interlock. Refer to Adjusting the cradle home table down interlock.