Doing a host Image Compute Node (ICN) time sync test
Purpose
This diagnostic tests the time synchronization between the host and ICN. It attempts to execute the date command on the ICN through the remote shell protocol. The ICN and host times are compared to make sure they meet specifications.
Components tested
- Remote shell service on the ICN
- Host and ICN time synchronization
Test sequence
- Select Host ICN Time Sync.
- Click Run.
- Make sure each ICN tested passes.
Results
| Result | Description | GE Field Action |
|
Passed |
The ICN has passed the diagnostic |
No further action required |
|
Failed |
The ICN Time Synch Diagnostic has failed |
|
|
Failed |
ICNs are not installed correctly |
|