- SIGNA MR355 / SIGNA MR360
- Service Manual
- 5856356-3EN Revision 5.0
- Basic Service Documentation. Copyright General Electric Company.
- 00000018WIA3077BF20GYZ
- id_131067791.2
- Jul 5, 2019 11:19:00 PM
Leak Sensor Location and Function
There are two Leak sensor routes.
Note:
Leak Sensor 1 is not used for Optima MR360/Brivo MR355 system
-
Leak Sensor 2
-
Leak Sensor 3
When leak sensor detects the leakage, cabinet monitor will react as follows.
-
If the Leak Sensor2 detects the leakage, LK2 LED is lit and error message is displayed on Host and Emergency Stop will work.
-
If the Leak Sensor3 detects the leakage, LK3 LED is lit and error message is displayed.