Finalizing Grafidy calibration

Procedure

  1. Exit the software tool, saving the eddy current data.
  2. Remove the Grafidy fixture from the system.
  3. If calibration is being run within ICW installation mode continue to the next calibration procedure. Otherwise, Save Grafidy eddy current data to SaveInfo media. See Saving information.