| GOC4 FRU Replacement |
Part Number |
Service Setup Required |
LFC Required |
Power-up Functional Passed |
System Retest Required |
Procedure Location |
| HP Computer |
2368697-3 |
Yes |
Yes |
Check for hardware present |
Scan & Recon 100 Image Series |
HP (Host) Replacement
|
| Software Installation Procedures Manual |
| Drives Swap |
n/a |
n/a |
n/a |
n/a |
n/a |
Optional Hard Drive Swap
|
| NIC Exchange |
n/a |
n/a |
n/a |
n/a |
n/a |
HP Host Computer NIC Exchange
|
| DARC-2 (Data Acq Recon Controller) |
- |
Yes |
Yes |
Check for hardware present |
Scan & Recon 100 Image Series |
DARC-2 Replacement
|
| Software Installation Procedures Manual |
| DARC2 Scan Disk Replacement |
2225004–9 |
See Procedure |
No |
See Procedure |
See Procedure |
DARC2 Scan Disk Replacement
|
| ICOM |
- |
Yes |
Yes |
Yes |
Scan & Recon 100 Image Series |
Intercom Board Replacement
|
| Software Installation Procedures Manual - reconfig only |
| IG (Image Generator) |
2368272 |
Yes |
n/a |
Yes |
Scan & Recon 100 Image Series |
GOC4 IG Node Replacement
|
|
Scan Images
|
| Media Converter 10base T to cat 5 |
2377227 |
n/a |
n/a |
Yes |
Scan & Recon 100 Image Series |
Media Converter Switch Settings
|
| DASM |
B7700L |
Yes |
n/a |
Yes |
Test Film |
n/a |
| Modem |
B7700MG |
Yes |
n/a |
Yes |
n/a |
n/a |
| Power Panel |
2174159-3 |
n/a |
n/a |
n/a |
n/a |
n/a |
| Rear Fan |
- |
- |
- |
- |
- |
GOC4 Fan Replacement
|
| Air Filter |
- |
- |
- |
- |
- |
- |