• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WHA30910CGYZ
  • Topic ID: id_2002361 Version: 9.2
  • Date: Oct 28, 2021 11:28:09 AM

Finalizing Grafidy calibration

Procedure

  1. Exit the software tool, saving the eddy current data.
  2. Remove the Grafidy fixture from the system.
  3. If applicable, remove the service filler panels from the table.
  4. If calibration is being run within ICW installation mode continue to the next calibration procedure. Otherwise, Save Grafidy eddy current data to SaveInfo media. See Running the Save and Restore tool.