• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA3037E010GYZ
  • Topic ID: id_2003172 Version: 7.1
  • Date: Jul 13, 2021 4:16:32 PM

Checking the body coil air flow

Tests and troubleshoots the function of the body coil air blower system which cools the patient bore wall surfaces with air forced between the gradient coil and the Radio Frequency (RF) coil.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-1 hour and 15 minutes-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.