Checks the function of the body coil air blower system.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 15 minutes | - |
| Required conditions |
|---|
| System software must be booted. |
Procedure
- Open Service Browser and select
- Select SRI Real-Time Data and click Run.
Result
Air Flow Sensor 0 = ##### 0.1 FPM should be displayed in the Test Results field.
- Confirm that the air flow feedback is greater than 440 fpm (= 4400 0.1 FPM in Test Results).
| Feedback displayed | | Result |
|---|
| Less than 440 fpm | | The system has failed the check. |
| Greater than or equal to the normal mode set point | | The system has passed the check and is functioning properly. |