- Discovery MR750 3.0T System Service Methods
- 5690009-2EN Revision 4
- 00000018WIA304D4840GYZ
- id_20149184.1
- Jul 12, 2021 1:02:58 PM
Installing the GOC hard drive - Dell T5820
Install the solid state drive (SSD) in the host computer.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 15 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Nonmagnetic Titanium Service Tool Kit, Small Set (can be used on magnets with magnetic fields > 1.5T) | 1 | 5113258 | - |
| ESD Mat | 1 | - | - |
| ESD Strap | 1 | - | - |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| 512GB SATA III SFF SSD 3D-TLS or V NAND Memory | 2 | 8780011-102 | - |
| Required conditions |
|---|
|
Use an antistatic wrist strap and mat while performing this procedure to prevent electrostatic discharge (ESD). Access to both the front and rear of the operator console is required. Move the console away from walls and remove obstacles in the room. |