Replacing the CMOS battery - Dell T5820
Removal and installation instructions for replacing the CMOS battery in the host computer.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 60 minutes | 10 minutes |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Screwdriver | 1 | - | - |
| Nonmagnetic Titanium Service Tool Kit, Small Set (can be used on magnets with magnetic fields > 1.5T) | 1 | 5113258 | - |
| ESD Mat | 1 | - | - |
| ESD Strap | 1 | - | - |
| DQA III Phantom | 1 | 2321556 | - |
| DQA Phantom Positioner | 1 | 5554497 | - |
| Consumables | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| GE Equipment Maintenance Sticker | 1 | 5661793 | - |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Dust Filter | 1 (if needed) | 5192248-5 | - |
| Adhesive-Backed Velcro, 2 x 4 inches | 2 (if needed) | 46-317475P2 | - |
| Host PC CMOS Battery, 3V, CR2032 | 1 | 5129534-3 | - |
| Required conditions |
|---|
|
Use an antistatic wrist strap and mat while performing this procedure to prevent electrostatic discharge (ESD). Access to both the front and rear of the operator console is required. Move the console away from walls and remove obstacles in the room. |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |