Discovery MR750w and SIGNA™ Architect T 3.0T System Service Methods
5690002-2EN
Revision
4
Object ID: 00000018WHA305C3EGYZ
Topic ID: id_2002769 Version: 4.0
Date: Nov 3, 2021 4:05:16 PM
Running LVShim
The objective of the LVShim tool is to fine-tune magnet homogeneity.
Prerequisites
Personnel requirements
Required persons
Preliminary requirements
Procedure
Finalization
1
-
30 minutes
10 minutes
Tools and test equipment
Item
Quantity
Part number
Manufacturer
LVShim Phantom Inner Sphere
1
2274476
-
LV Phantom Small Half Assembly
1
2306642
-
LV Phantom Big Half Assembly
1
2306645
-
Curved Adapter Panel for Flat Table (GEM)
1
5395828
-
Universal Phantom Pad
1
46-328383P1
-
Service Filler Panel
1
5344577
-
Note: The LV Phantom Small Half Assembly, LV Phantom Big Half Assembly, and LVShim Phantom Inner Sphere are used for systems running software versions DV26.0 or earlier. The LVShim Phantom Inner Sphere alone is used with the Universal Phantom Pad for systems running software versions DV29.1 or later.
Required conditions
Confirm that passive shimming is complete and is within specification.
Remove all coils from the cradle and phantoms from the bore.
Safety
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
Comply with all site-specific training and workplace safety requirements.
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.
Warning
Poisonous substance
Green phantoms contain nickel, a suspected carcinogen.
Do not ingest. Dispose of hazardous waste according to state and federal regulations.
Warning
Poisonous substance
Pink phantoms contain silicon oil, a potentially poisonous substance.
Do not ingest. Dispose of hazardous waste according to state and federal regulations.
Note: If the system is running DV29.1 or later, no side lobes are required for shimming. The system will report an error if the side lobes are used. If the system is using DV26.0 or earlier, side lobes are required for shimming.