Discovery MR750w and SIGNA™ Architect T 3.0T System Service Methods
5690002-2EN
Revision
4
Object ID: 00000018WIA30A5A650GYZ
Topic ID: id_2024933 Version: 2.5
Date: Nov 8, 2021 2:09:22 PM
Running LVShim - Gradshim for systems running DV29.1 or later
Procedure
(For curved tables) Install a service filler panel under both cradle top panels.
Figure 1. Installing service filler panels on curved table
Set the universal phantom pad on the table and align the crosshairs at 1080 mm away from home.
Put the sphere on the pad with the strap (in the groove) in the orientation shown below, making sure the laser lights are cross-centered on the sphere and phantom pad.
Figure 2. Sphere phantom position
Note: This image is a representative example. Actual systems may vary.
Set up and align the LVShim phantom. Align the phantom so that:
The strap on the phantom is aligned parallel to the table surface.
The phantom is centered left to right with respect to the cradle top.
Landmark on the universal phantom pad crosshairs.
Note: The landmarking tolerances of the LVShim tool are: I/S: ± 3 mm, A/P: ± 30 mm, R/L: ± 5 mm. Ensure accuracy of the placement for landmarking.
Note: Due to the tight landmarking tolerance, it is important to align the lasers on the cross lines of the universal positioner before setting the sphere.
Press Advance to Scan.
Note: The tool checks phantom positioning and will fail if the above landmarking specs are out of tolerance.
Start the LVShim tool from the Common Service Desktop on the Calibration tab.