- Discovery MR750w and SIGNA™ Architect T 3.0T System Service Methods
- 5690002-2EN Revision 4
- 00000018WIA3026CE20GYZ
- id_131061683.3
- Nov 5, 2020 1:38:56 PM
Probe Tuning and SNR Check Tool (DV24 and later)
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | 5 minutes | 45 minutes | Not Applicable |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| MRS Phantom (without loader) | 1 | - |
2152220 | - |
| 3.0T Split-Top Head Coil | 1 | - |
5182872 | - |
| Curved Adapter Panel for Flat Table (GEM) | 1 | Flat (GEM) table |
5395828 | - |
| Service Filler Panels | 2 | Curved table |
5344577 | - |
| ||||
| Condition | Reference | Effectivity |
|---|---|---|
|
PROBE-S option (M3333WH) is required at site to perform PROBE-S Cal & SNR Check or PROBE-P Cal & SNR Check. | - | - |
|
PROBE-P option (M3333WG) is required at site to perform PROBE-P Cal & SNR Check or PROBE-P SNR Check. | - | - |
About this task
This tool is available in systems with DV24 and later software.
The Probe Tuning and SNR Check Tool is an essential part of conditioning a system for performing spectroscopy, to optimize both SNR and phasing. All tests can be run in one session, and no interaction is needed during the calibration.
There are two Probe/SV pulse sequences: Probe-s and Probe-p.
-
The Probe-s sequence is a version of the stimulated echo acquisition mode (STEAM). This sequence is used for short echo time, single-voxel spectroscopy; that is, TEs shorter than 35 milliseconds.
-
The Probe-p sequence is a version of the point-resolved spectroscopy (PRESS) pulse sequence. The PRESS sequence provides twice the SNR of the STEAM sequence but at the expense of slightly longer minimum TEs.
This document describes two separate hydrogen only spectroscopy (PROBE/SV) procedures:
-
Echo Peak Location calibration (automated PROBE tuning script)
-
PROBE signal-to-noise ratio (SNR) performance test
Scan Setup
About this task
This section provides instructions for centering the phantom. There is no need to run a 3-plane localizer before scanning, because the tool automatically performs a phantom centering check. The tool also has a built in wait time to allow the phantom to settle before scanning.
Procedure
Reviewing Results
About this task
The probeTune tool also generates log files, which can be viewed from the tool or manually. There are four log files (one for each probe calibration and SNR check). The log files are stored in /usr/g/service/data and are named to indicate which test they result from (for example, *****_CAL.PRB and *****_SNR.PRB). There is no need for data sheets as in previous probe tuning applications. The files are always on the system, and saved on the SaveInfo media (if SaveInfo is performed after calibrations).
Procedure
Finalization
Finalization
-
If not performing this test as part of an installation, perform a SaveInfo to save the calibration file.
-
Remove all service equipment from the system, including filler panels if applicable.




