Signal To Noise Check-Body Scan
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | - | - | - |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| · Body SNR Sphere | 1 | - |
46-265635G4 | - |
| · SPT Body Loader or Long Body Loader | - | - |
2135652-2 46-287902G1 | - |
| ||||
| Condition | Reference | Effectivity |
|---|---|---|
|
No image artifacts | - | - |
|
System Gain Calibration complete | - | - |
|
System Gain Calibration complete | - | - |
About this task
Note:
The Body SNR Sphere Phantom, 46-265635G4, is not provided with the System. If the Body SNR sphere is needed, the customer must order it. A Body Loader is provided with the system.
Procedure
Finalization
No finalization steps.



(Image Browser) icon. When the browser is displayed, click 

(scan) icon. In the SCANNING RANGE area, enter that value for
the A/P Start and End values (anterior offset) of the coronal scan.