InSite Remote Service Platform (RSvP) pre-installation requirements

Describes the configuration and information required prior to the installation of the Remote Service Platform (RSvP).

RSvP pre-installation requirements

The MR scanner requires connectivity directly to the Internet.
  • The MR system allows for DNS configuration or proxy server-based connection to the Internet.
  • A current Internet connection supporting InSite1 connection can be reused.
  • RSvP uses SSL/TLS protocol, https over port 443.
  • In rare cases when the customer does not accept the above connection protocol, after the issue is escalated, the local connectivity team can provide help to connect through SSL/TLS proxy IP over the site-to-site VPN.
The following information is required from the customer prior to starting an upgrade to RSvP:
  • Final system ID for the site (final registration in FFA).
  • DNS IP address or proxy IP address (required for modality configuration / Internet access).
  • If a customer wants to only whitelist the specific URLs, the following are the required URLs for RSvP connectivity:
    • https://insite.gehealthcare.com
    • https://as1-insite.gehealthcare.com
    • https://as2-insite.gehealthcare.com
    • https://gehealthcare-ns.flexnetoperations.com
    • https://download.flexnetoperations.com

If the software being upgraded to RSvP is already connected through InSite1, do Checking Internet connectivity prior to InSite Remote Service Platform (RSvP) upgrade to confirm the customer's Internet connection before the software is upgraded to RSvP.