- id_15667663
- Version: 2.2
- Date: Feb 18, 2020 10:15:14 AM
Doing RF calibration datapath diagnostic (DPP)
The RF calibration datapath diagnostic tests the following calibrations: Transmit Gain (TG) calibration for the narrowband exciter (DTX-1) and (if installed) the broadband exciter (DTX-2); DC offset adjustments for the RRX modules; and R1 gain measurements for the receiver modules.
Disclaimer
Depending on your service agreement, not all service tools, diagnostics, and utilities referenced in this document may be accessible. Contact your sales person for information on available service license packages.