- id_12374319
- Version: 1.1
- Date: Jul 5, 2019 10:03:33 PM
Manual Entry Probe/SV SNR Protocols
This document contains the Non-Proprietary protocols for Testing Probe-S and Probe-P SNR. Refer to Probe/SV Calibration and SNR Tests for the complete calibration procedure, and substitute the protocols below for the Service protocols listed.
-
Refer to Table 1 for the non-proprietary Probe-S SNR Protocol.
-
Refer to Table 2 for the non-proprietary Probe-P SNR Protocol.
|
Patient Register New Pt PATIENT INFORMATION Patient ID geservice Patient Name probes snr Weight (Lb) 111 (IMPORTANT) Landmark Landmark >Nasion Patient Protocols Patient Position Patient Position Patient Position >Supine Patient Entry >Head First Coil ...HeadAccept Imaging Parameters Plane >Axial Mode >MRS Pulse Seq ...Probe-SAccept Imaging Options Ext. Dyn. Range Psd Name (no entry) Protocol (no entry) Scan Timing # of Echoes 1 (default) TE 30 TR 2000 |
User CVs Screen CV3 scan mode1.00 CV4 total # of scans32.00 Scanning Range FOV24 Start S10 EndI10 StartR10 EndL10 StartA10 EndP10 Table Delta0.00 (default) Acquisition Timing Freq1 Phase1 NEX2.00 Freq Dir>A/P Auto Center Freq>Water, Autoshimselected, Contrastnot selected Save series Prepare to Scan Research Operations Display CVs CV Name CV Value tempC (greenest value on MRS Phantom) Accept Research Operations Download |
|
Patient Register New Pt PATIENT INFORMATION Patient ID geservice Patient Name probep snr Weight (Lb) 111 (IMPORTANT) Landmark Landmark>Nasion Patient Protocols Patient Position Patient Position Patient Position>Supine Patient Entry>Head First Coil...HeadAccept Imaging Parameters Plane>Axial Mode>MRS Pulse Seq...Probe-PAccept Imaging Options Ext. Dyn. Range Psd Name (no entry) Protocol (no entry) Scan Timing # of Echoes 1 (default) TE 35 TR 2000 |
User CVs Screen CV3 scan mode 1.00 CV4 total # of scans 32.00 CV15 Phantom = tempC value CV17 = 0 CV18 = 7 Scanning Range # CSI Slices1 FOV24 Start S10 EndI10 StartR10 EndL10 StartA10 EndP10 Table Delta0.00 (default) Acquisition Timing Freq1 Phase1 NEX2.00 Freq Dir>A/P Auto Center Freq>Water, Autoshimselected, Contrastnot selected Save series Prepare to Scan Research Operations Display CVs CV Name CV Value tempC (greenest value on MRS Phantom) Accept Research Operations Download |