Procedure
- Alert necessary personnel that power is being restored.
- If LOTO device(s) were installed, remove personal LOTO device(s) from both return and supply valves.
- If return and supply valves were shut off, return the handles in-line to the 0º angle. Check for coolant leaks.
- If LOTO device was installed, remove personal LOTO device from the HEC circuit breaker on the MDP.
- If the MDP circuit breaker was turned off at the front panel of the MDP, turn on the HEC circuit breaker.
Finalization
- After the pumps turn on, check for coolant leaks.
- Take the Magnet Monitor out of Service mode.