• Object ID: 00000018WIA30CD4E40GYZ
  • Topic ID: id_2019795 Version: 2.2
  • Date: Apr 23, 2020 8:48:07 PM

Installing the Waveguide sub-panel

Prerequisites

The original waveguide sub-panel must be removed and replaced with the custom waveguide to allow cabling into the magnet room.

Procedure

  1. Perform LOTO on the PGR and penetration cabinets .
  2. Perform LOTO on the Heat Exchanger Cabinet (HEC) to remove power from the cryocooler compressor, which is necessary to complete this procedure.
  3. Install the custom waveguide panel (2394034-10) in place of the standard panel.
    Figure 1. Sub-panel
    Note: If the waveguide panel has preexisting (custom) modifications, they must be reapplied to the sub-panel before BrainWave can be installed.