Patient handling adjustments
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | Not applicable minutes | 300 minutes | Not applicable minutes |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Non-ferrous standard head screwdriver | 1 | - | - | - |
About this task
Overview
The proper alignment of the entire Patient Handling System is important for good image quality output from the GE MR Scanners.
Procedure
Perform table adjustements as required:
- Bridge alignment
- Dock centering and dock hook tension adjustment
- Table top height adjustment
- Cradle release pin height adjustment
- Patient table movement verification
- Checking the longitudinal drive belt tension
- Leveling patient table
- Cradle/LPCA release adjustment
- Cradle/table release block adjustment/cradle egress functionality