Day 2 |
INSTALLER #1
|
|
|
STEP E0: Mechanical-Electrical Transition
|
|
8:00 |
STEP E1: Inventory System Software and Restore
System State
|
| 9:00 |
STEP E2: System and Console Configuration
|
|
10:30 |
STEP E3: Install Software Options and Complete
Computer Integration
|
|
12:00 |
LUNCH BREAK |
|
1:00 |
STEP E4: Autovoice/Intercom Checks
|
|
1:15 |
STEP E5: Alignment Light & Table Height
Characterization
|
|
2:15 |
STEP E6: Gantry/Table Interference Test
|
|
2:45 |
STEP E7: Tube Warm Up and Fast Calibration
|
|
3:00 |
STEP E8: Table/Gantry Alignment
|
|
4:00 |
STEP E9: Center Phantom
|
|
4:30 |
STEP E10: Tomographic Plane Indication
|
|
Day 3 |
INSTALLER #1 |
|
8:00 |
STEP E11: Image Quality Test
|
|
10:00 |
STEP E12: Verify Network Connections
|
|
10:15 |
STEP E13: Install and Verify Customer Imaging
Options
|
|
12:15 |
LUNCH BREAK |
|
1:15 |
STEP E14: System Functional Test
|
|
2:45 |
STEP E15: Administrator Account Configuration
in EA3 and Save System State
|
|
3:30 |
STEP E16: Patient Touch and System Chassis Leakage
Tests
|
|
4:30 |
STEP E17: Install Service Cabinet
|
|
Day 4 |
INSTALLER #1 |
|
8:00 |
STEP E18: Install Covers
|
|
10:00 |
STEP E19: X-Ray On Indicators, Warning Labels,
and Product Locator Cards
|
|
10:30 |
STEP E20: GE and Regulatory Forms
|
|
11:30 |
STEP E21: Final Installation Checklist
|