Checking the GOC battery replacement
Check the battery replacement sticker on the GOC.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 5 minutes | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
Procedure
Check the battery replacement sticker on the GOC.
The sticker should be attached to lower left or right side of the GOC. Batteries should be replaced every three years.
- If the date on the battery replacement sticker is within 12 months, order the battery and replace it during the next PM. See .
- If the GOC does not have a sticker, order a battery and replace it during the next PM. See .