The objective of the LVShim tool is to fine-tune magnet homogeneity.
Prerequisites
Personnel requirements
Required persons
Preliminary requirements
Procedure
Finalization
1
-
30 minutes
10 minutes
Tools and test equipment
Item
Quantity
Part number
Manufacturer
LVShim Phantom Inner Sphere
1
2274476
-
LV Phantom Small Half Assembly
1
2306642
-
LV Phantom Big Half Assembly
1
2306645
-
Safety
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
Comply with all site-specific training and workplace safety requirements.
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.
Warning
Poisonous substance
Green phantoms contain nickel, a suspected carcinogen.
Do not ingest. Dispose of hazardous waste according to state and federal regulations.
Warning
Poisonous substance
Pink phantoms contain silicon oil, a potentially poisonous substance.
Do not ingest. Dispose of hazardous waste according to state and federal regulations.