• Discovery MR750 3.0T System Service Methods
  • 5690009-2EN Revision 4
  • Object ID: 00000018WIA306E0110GYZ
  • Topic ID: id_2003328 Version: 11.8
  • Date: Dec 9, 2021 7:18:58 PM

Doing the System Performance Test (SPT)

The system performance test (SPT) automatically executes Eddy Current Check, SNR, Stability, and Coherent Noise tests. To do SPT in PM mode, you must launch SPT from the PM Assist menu.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-90 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Head Loader Positioner15110241-
Head TLT Loader for 3.0T12360031-
Head TLT Sphere for 3.0T (Pink Solution)12359877-
Body TLT Sphere for 3.0T (Pink Solution)12360025-
Body TLT Loader for 3.0T12360037-

Service Key

1--
Required conditions
Cancel all previous exams. Click End Exam.
Remove any patient positioning devices, such as patient restraints or other attached devices.
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

About this task

The next PM date reporting and PM due/overdue notices were removed from SW versions 27 and later. Please reference the CRM for this customer/system ID to ascertain the next PM due date.