• Discovery MR750 3.0T System Service Methods
  • 5690009-2EN Revision 4
  • Object ID: 00000018WIA307FE010GYZ
  • Topic ID: id_2003262 Version: 2.38
  • Date: Mar 3, 2021 6:21:08 PM

Calibrating High Order Shim (HOS)

This procedure creates High Order Shim (HOS) reference maps characterizing the fields created by the HOS windings. These maps are used by the HOS feature.

Prerequisites

Personnel Requirements
Required personsPreliminary requirementsProcedureFinalization
1-60 minutes -
Tools and test equipment
ItemQuantityPart numberManufacturer
LVShim Phantom Inner Sphere12274476-
Universal Phantom Pad146-328383P1-
Note: For LVShim procedures, continue to use side lobes.
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.