- Discovery MR750 3.0T System Service Methods
- 5690009-2EN Revision 4
- 00000018WIA30DD3E20GYZ
- id_131062998.0
- Feb 21, 2021 9:07:47 PM
Doing Multi-Nuclear Spectroscopy (MNS) functional checks
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | 15 minutes | 45 minutes | 30 minutes |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| 13C 3.0T | 1 |
2354050-6 - T/R Module 5114403-3 - QA Probe | - |
| 13P 3.0T | 1 |
2354050 - T/R Module 5114403-3 - QA Probe | - |
| 9-hole USS and Grafidy base plate | 1 | 46-271410 | - |
| Inner DC Block | 1 |
5309801 | - |
| Required conditions | |
|---|---|
| Condition | Reference |
| MNS option installed and amplifier calibrated. | - |
| Each time the QA coil is hooked up, the Inner DC block must be connected onto BNC connector of the QD box. | - |
Note: (For PET/MR systems) Remove the CMA coil and replace it with the CMA proxy. MNS will not operate with the CMA coil.
About this task
| Nucleus | Field Strength | Min Mag at TG = 0 | Min. SNR | Max NoiseStdev |
| 13C |
3.0T | 2.0 X 10^8 | 120 | 60000 |
| 31P | 3.0T | 2.0 X 10^8 | 280 | 60000 |

The following procedures are included in this functional test:
- See Functional test for the Functional Test.
- See Signal to noise test for the Signal to Noise Test.
- See Noise test for the Noise Test.
Functional test
Hardware Setup
Procedure
Scanner setup
Procedure
Signal to noise test
About this task
Hardware setup
Procedure
- Install the T/R module onto the top of the LPCA cover.
- Place the grafidy base plate on the patient bed near the top of the bed.
- Place the QA probe in the center hole of the grafidy base plate and lock the probe into place.
- Connect Inner DC Block to BNC connector. See Figure 3.
- Connect the QA probe cable to the quick disconnect module.
- Connect the QD module to the T/R module.
- Landmark on the center of the QA probe and advance to scan.
Scanner setup
Procedure
Scanning and analysis
Procedure
Noise test
About this task
If the Functional Test procedure, Functional test, has successfully completed and you now want to begin the Noise Test, select Noise Scan from the MNS Setup area and select Setup. Then proceed to Step 9.
Hardware setup
Procedure
- Install the TR module onto the top of the LPCA cover.
- Connect the cables to the A connector on the LPCA.
- Place the grafidy base plate on the patient bed near the top of the bed.
- Place the QA probe in the center hole of the grafidy base plate and lock the probe into place.
- Connect Inner DC Block between the QA probe cable and the quick disconnect module. See Figure 3.
- Connect the QD module to the TR module.
- Landmark on the center of the QA probe and advance to scan.
Scanner setup
Procedure
Scanning and analysis
Procedure
Finalization
Procedure
- Remove any hardware used during system testing.
- Do a test scan of the system to make sure it is operating correctly.

















