• Discovery MR750 3.0T System Service Methods
  • 5690009-2EN Revision 4
  • Object ID: 00000018WIA30501450GYZ
  • Topic ID: id_2021715 Version: 2.2
  • Date: Oct 11, 2021 6:20:13 PM

Removing the MNS exciter

Procedure

  1. Perform LOTO on the RF amplifier and the PEN cabinet (magnet room electronics). See the MR Service Safety Manual, PN 5452735.
  2. Remove the thumbscrews that secure the MNS exciter to the PEN cabinet.
    Figure 1. MNS Exciter Cabinet
    ACAN fiber-optic board (CFB)
    BNB exciter board
    CMNS BB exciter board
    DThumbscrews securing module to cabinet frame
  3. Disconnect all the cables to the MNS exciter. Be sure to note cable locations.
  4. Carefully slide the MNS exciter out of the PEN cabinet.