- SIGNA MR355 / SIGNA MR360
- Service Manual
- 5856356-3EN Revision 5.0
- Basic Service Documentation. Copyright General Electric Company.
- 00000018WIA30098F20GYZ
- id_131067502.0
- Jul 19, 2019 11:00:46 AM
UPM- Body and Head Functional Check
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | Not Applicable | 15 minutes | 5 minutes |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Power Measurement Kit | 1 | - |
46-317724g1 or g2 | - |
| 25 KW 30 dB Attenuator (Dummy Load) | 1 | - |
46317724p14 | - |
| 50 Ohm Terminator | 1 | - |
46-265874p1 | - |
| 7/16Male -N Female Adapter (Attached in System Cabinet) | 1 | - |
5166139 | - |
| ||||
| Condition | Reference | Effectivity |
|---|---|---|
|
Properly calibrated RF Head, Body . | - | - |
About this task
The UPM Functional checks for 1.5T consist of three automated tests:
-
Peak Power, Pulse Width, Duty Cycle at TG=170
-
RF Inhibit test (Check that RF is disabled when inhibit condition)
-
Trip Test
All 3 automated tests must pass to ensure the correct operation of the Universal Power monitor system.
Run both Head and Body UPM Functional checks for the proper system type.
1.5T BODY UPM FUNCTIONAL CHECK
1.5T BODY UPM FUNCTIONAL CHECK Setup
Procedure
Setting up Software Tool to Calibrate UPM – Body Forward Power
About this task
This tool must see a Landmark only. It is not necessary to have a Body Phantom in place. Simply Landmark on the cradle where the Body Phantom would be.
Procedure
1.5T HEAD UPM FUNCTIONAL CHECK
1.5T HEAD UPM FUNCTIONAL CHECK Setup
Procedure
Setting up Software Tool to Calibrate UPM – Head Forward Power
About this task
It is necessary to have a Head coil in place. It is not necessary to have a Head Phantom in place. Simply Landmark on the Head Coil.
Procedure
Troubleshooting UPM
Procedure
UPM Theory
About this task
The Universal Power Monitor calibration parameters found in the UPM Calibration file (/export/home/signa/tools/upm.config) are used in the calibration/Functional Check program. The Program code sets the Digital attenuators in the UPM based on the results of the 3 tests run, and a comparison to these UPM Calibration parameters. This calibration file is accessible from within the UPM Calibration tool from the “File” pull-down Menu, “Edit Config” Option. These values can be edited and saved during troubleshooting to perhaps understand where the power monitor is tripping, (If at all). Before changing any parameter, Make sure to select and run the “Bkup UPM Cal” option first. This will allow you to return to the default values at any time. The Screen shot of the UPM Calibration file found in this document contains the default configuration values for all of the UPM parameters as of December 8, 2003. (Use the [Restore Defaults] on the config file editor GUI to always get back to known default values).The final UPM digital attenuator values are written to the UPM Calibration Configuration files (UpmCal1.cfg, UpmCal2.cfg.etc.,) located in directory: w/config

















