• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30498F20GYZ
  • Topic ID: id_13107595 Version: 2.0
  • Date: Jul 19, 2019 11:00:47 AM

Troubleshooting for UPM- Body and Head Functional Check for HDE14.0

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable30 minutes5 minutes
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
50 Ohm Terminator1-

46-265874p1

-
7/16Male -N Female Adapter (Attached in System Cabinet)1-

5166139

-
Table 3. Safety
Table 4. Required conditions
ConditionReferenceEffectivity

Properly calibrated RF Head, Body .

--

This troubleshooting needs to be performed if the failure of UPM Functional Check is related to Over limit or Under limit of peak power.

Finalization

  1. Check that cable at SRFD is restored
  2. Check that SW2 of Driver Module is set to enable.
  3. Return the system to patient scanning configuration and do a test scan.