Basic Service Documentation. Copyright General Electric Company.
Object ID: 00000018WIA3049CF20GYZ
Topic ID: id_13106837 Version: 1.2
Date: Jul 5, 2019 10:46:04 PM
SNR Image Analysis
Prerequisites
Table 1. Personnel requirements
Required persons
Preliminary requirements
Procedure
Finalization
1
-
-
-
About this task
The SNR tool retrieves two operator-selected images. Signal
value is computed as the mean pixel value in an ROI covering 80% of
the image. The image is analyzed to determine the center of the image
for positioning the ROI. Subtracting the second image from the first
creates a difference image and the same ROI is used to calculate noise
from the subtracted image. The signal value, noise value, and signal-to-noise
ratio are reported. There is an option to save the difference image
with the results annotated.
Procedure
On the Service Desktop Manager, click the Service Browser button.
Follow the instructions on Figure 1, below.
Figure 1. OPENING THE IMAGE QUALITY NEMA TOOL FROM THE COMMON SERVICE
DESKTOP
For TwinSpeed, highlight the same GradMode as was used in section
2, 3 or 4 above, then click on OK.
A NEMA Image Quality window will appear on the desktop. Type 1 and press Enter for the Signal
to Noise Check. See Figure 2.
Figure 2. NEMA IMAGE QUALITY MENU
Refer to Illustration 5-3 for the next steps.
Figure 3. NEMA IMAGE SELECTION MENU
At the prompt, type A (Select Exam) and
then enter the exam number.
At the prompt, type B (Select Series) and
enter the series number.
Type C (Select Image) and enter the image
number.
Type X (Execute the selected test) to run
the procedure.
You’ll be prompted “Choose Second Image.”
Continue by doing the following:
Type A and then enter the exam number.
Type B and enter the series number.
Type C and enter the image number.
Type X to run the procedure.
The analysis then begins. The final values are displayed on
the screen, as shown in Figure 4.
Figure 4. SNR RESULTS SCREEN
Record Signal, Noise, and Signal-to-Noise in the Data Sheet at the end of this document.
Repeat steps Step 5 through
12 above, selecting the next exam for the analysis of each remaining
image pair.
Keep the completed data sheets with the system data sheets for
future reference.
Type 6 and press Enter to exit the NEMA Image Quality Analysis Menu.