Note: Updates should not be loaded with application software running.
Procedure
Either insert the update CD/DVD in the appropriate drive and select CD/DVD, or select Repository to load updates from /usr/g/insite/data.
Figure 1. Guided Install patches
Click on the applicable update to load and select Install.
Note: You must load one update at a time.
Figure 2. Available updates
Result
Updates that load successfully display in the Installed Patches window. A confirmation message also appears in the GUI status section.Figure 3. Install done message
The update process looks for the software revision and conflicting information before loading. If the software revision of the update does not match the applications revision, a message displays in the status section of the GUI that explains that the update cannot be loaded.
If the message Patch_tst_M4_01 = PA conflicts with patch_tst_M4_03-PA-1 appears, you must delete the most recently installed update from the system before you load the new update. See Removing software updates.
Perform any other patch load instructions that are documented in the “Read Me” file of the Service Pack.
Reboot the system. When prompted, log in as sdc for user.
Finalization
Perform a check scan to verify that the system is functioning properly. See Doing a check scan