Note:
Normally, the voltage of upper terminal will be 0 V after
30 minutes.
For lower terminal, the voltage may exist in case of switch
failure in XFD-PS and the charge of super capacitor cannot be discharged.
If voltage exists in lower terminal, please check that SW1 of
XFD-PS is turned to discharge position.
If SW1 is at discharge position, waited for 30 minutes, and
voltage still exists at lower terminal, wait until the voltage becomes
0V.
While waiting, please install cover for XFD Power Pannel Assy.
At worst case, it will take 1 days to discharge.