Geometry Verification-Software Configuration Check

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable10 minutesNot Applicable

About this task

The system configuration values for the Setting on the x-, y-, and z-gradient to obtain 1 gauss/cm (also known as cfxfull, cfyfull, and cfzfull, respectively) should be either all positive, or all negative, depending on which direction the magnetic field is ramped-up. If the magnet was ramped with the B0 field going into the magnet (usual case for 1.0T, 1.5T and 3.0T systems), the cfxfull, cfyfull, and cfzfull values in the configuration file must be positive in order for the geometry to come out right.

Last Update09/28/2005

Procedure

  1. From the Common Service Desktop, Select Configuration, Config File Manager and launch the tool. See Figure 1
    Figure 1. Opening Config File Manager From Service Desktop
  2. On the Config File Manager main screen, Click the Gradient Config File WHOLE button on the top.
  3. Verify that all three gradient values are the same sign (all positive or all negative).
  4. When finished editing values Click Save Changes and then Quit button to exit application.
  5. Click Yes in the dialog box question, “Really Quit?”
  6. Click Gradient Config File Zoom and repeat Step 3.
  7. Next, a Save dialog box appears for each configuration file that has changed. To save changes click the Save button. To not save changes click the Do Not Save button.

Finalization

  1. Exit Configuration File Manager
  2. Continue on to verifying Gradient Polarity Checks or Return to the Geometry Verification main menu